Turbidity: Self-Cleaning
Self-Cleaning Turbidity Sensor

Hydrolab's Self-Cleaning Turbidity sensor measures from 0 to 3000 NTU and includes a user-programmable cleaning system to remove any fouling or debris that could otherwise affect readings.

 

Features:

  • ISO 7027 compliant
  • User-programmable self-cleaning system can perform up to 10 cleaning cycles before each reading
  • Accurately measures up to 3000 NTU

Benefits:

  • Fixed parking position ensures consistent data collection after each cleaning cycle
  • 3000 NTU range allows Turbidity tracking even during rain storms or other events that could cause abnormally high readings
  • Exceptional linearity even in high NTU environments
  • Utilizes small aperture technique to reduce false readings from particulates and other debris

Product Data:

Specs  
Range: 0-3000 NTU
Accuracy (compared to StablCal):  ± 1% up to 100 NTU,
± 3% from 100-400 NTU
± 5% from 400-3000 NTU
Resolution: 0.1 NTU from 0-400 NTU;
1 NTU for >400 NTU
Temperature Coefficient: 0.05%/C
Sensor housing: Stainless steel:

Standard housing for typical fresh water applications and depths to 200 M.

  Plastic:

Corrosion-resistant housing for use in aggressive saline environments such as oceans, bays and estuaries. Rated to depths of 50 M.

 

  Consult with your sales representative to determine which sensor best fits your needs.
     
 
Hach Company Privacy Policy Site Map